5-16
ES User’s Guide
Optimizing Measurements
Reducing Mismatch Errors
Reducing Mismatch Errors in a Transmission
Measurement
The best way to reduce mismatch errors in a transmission measurement
of a two-port device is to perform a two-port calibration directly at the
point that will connect to the DUT, using the exact frequency parameters
that you will be using for the measurement. Though not quite as
accurate as a two-port calibration, an enhanced response calibration can
also be used to reduce source match errors. To minimize load match
error, it is recommended that you use an attenuator on the output of the
DUT. An enhanced response calibration offers the advantage of faster
measurement speeds than when using a two-port calibration.
1. Set up your S
21
forward transmission measurement or S
12
reverse
transmission measurement, and frequency parameters.
2. Press
or
.
3. Follow the instructions on the display.
See
for more information on calibration choices for
transmission measurements.
NOTE
If you choose to do an enhanced response calibration, it is recommended
that you use a high quality attenuator on the output of the DUT, as
shown in
. Six to ten dB of attenuation should be
sufficient to significantly reduce mismatch errors.
NOTE
Using an attenuator on the output of the DUT will reduce the system
dynamic range.
CAL
User 2-Port
CAL
User Response
Enhanced Response
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...