2-10
ES User’s Guide
Getting Started
Performing the Operator's Check
Performing the Operator's Check
The operator's check should be performed when you receive your
instrument, and any time you wish to have confidence that the analyzer
is working properly. The operator's check does not verify performance to
specifications, but should give you a high degree of confidence that the
instrument is performing properly if it passes.
The operator's check consists of making the following measurements
with the cable that was supplied with your analyzer:
• S
21
and S
12
• broadband power
• S
11
and S
22
• S
11
and S
22
(with a 50
Ω
or 75
Ω
load, instead of the cable)
Equipment List
To perform the operator's check you will need the following:
• A known good cable such as the one that was supplied with your
analyzer. The cable you use should have
≤
0.5 dB of insertion loss up
to 1.3 GHz and
≤
0.75 dB of insertion loss from 1.3 to 3.0 GHz.
• A known good load (> 40 dB return loss) that matches the test port
impedance of your analyzer such as one from calibration kit
HP 85032B/E (50
Ω
) or HP 85036B/E (75
Ω
).
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...