ES User’s Guide
3-35
Making Measurements
Measuring S-Parameters using a Two-Port Calibration
EXAMPLE #2: the noninsertable device has a 3.5 mm(f) input
connector, and a type-N(f) output connector.
NOTE
For analyzers with 50 ohm input impedance only: in the 50 ohm type-N
calibration kit (HP 85032B), select the APC-7-to-type-N(f) adapter and in
the 3.5 mm calibration kit (HP 85033D), select the APC-7-to-3.5 mm(f)
adapter. Connect these two adapters to create the type-N-to-3.5 mm
female-to-female adapter in the following example.
1. Connect a test cable with 3.5 mm(m) connectors to the analyzer’s
port 1. (This example assumes a type-N(m)-to-3.5 mm(f) adapter is
connected directly to the analyzer’s port 1.)
2. Connect a test cable with type-N(m) connectors to the analyzer’s
port 2.
3. Press
(
or
)
.
4. Select the proper calibration kits for each port:
• For port 1, use the front panel knob or the
keys to
highlight Port 1 in the display’s table. Press
.
• For port 2, use the front panel knob or the
keys to
highlight Port 2 in the display’s table. Press
.
5. Press
. Perform a one-port reflection
calibration using type-N(f) standards on the type-N(m) open-end of
the port 2 test cable. Refer to
“Perform a User One-Port Calibration”
.
6. Connect the type-N-to-3.5 mm female-to-female adapter to the
type-N(m) open-end of the port 2 test cable.
7. Add a 3.5 mm short to the open end of the adapter as shown in
PRESET
User Preset
Factory Preset
CAL
More Cal Cal Kit
3.5 mm
Type-N(m)
MEAS 1
S22 Refl Port2
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...