8-28
ES User’s Guide
Hardkey/Softkey Reference
E
Softkey in the max line and min line menus. Sets the end (or stop)
amplitude (height) of a limit line.
See
“Using Limit Testing” on page 4-28
lines.
Access Keys:
,
or
Softkey in the secure menu that is used with the IBASIC feature of the
instrument.
See the HP Instrument BASIC User's Handbook for information on the
secure function.
Access Keys:
Softkey in the response cal menu when in S
21
or S
12
(transmission)
measurement mode. Select this key to perform a user-defined
enhanced response calibration. Restores a previous enhanced response
calibration if current analyzer parameters are compatible with those
used during the previous calibration. Corrects for frequency response
and source match errors.
See
“Measuring S21 Forward Transmission using an Enhanced
Response Calibration” on page 3-41
Access Keys:
Softkey inthe title and clock menu. Displays the menu to edit the
screen title on line 1 of the display. Maximum number of characters is
36.
Access Keys:
Softkey in the title and clock menu. Displays the menu to edit the
screen title on line 2 of the display. Maximum number of characters is
36.
Access Keys:
End Limit
DISPLAY
Limit Menu Add Limit
Add Max Line
Add Min Line
End Line #
SYSTEM OPTIONS
IBASIC Utilities Secure
Enhanced
Response
CAL
User Response
Enter Line 1
DISPLAY
More Display Title and Clock
Enter Line 2
DISPLAY
More Display Title and Clock
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...