4-24
ES User’s Guide
Using Instrument Functions
Using Markers
To Use Delta (
∆
) Marker Mode
In delta marker mode, a reference marker is placed at the active marker
position. All marker values are then displayed in reference to this delta
marker. When the amplitude of the measurement trace changes, the
reference marker value also changes. The delta marker is represented by
a triangle pointing up (
∆
) with a delta symbol (
∆
) below it rather than a
number.
Limit testing can be performed on delta amplitude and delta frequency
using specific markers. See
“To Use Marker Limit Functions” on
chapter.
To follow along with this example, set up the instrument for an S
21
forward transmission measurement of the bandpass filter that was
shipped with your instrument.
1. Press
to move marker 1 (the active marker) to the maximum
value on the trace.
2. Press
to place a reference marker at the active marker
position.
3. Press
to place marker 1 to the minimum point on the trace.
4. The difference between the markers' frequency and amplitude value
is shown in the upper-right corner of the analyzer screen.
5. Press
. Use the front panel knob to
move marker 2 towards the right-hand side of the display screen.
Note that the marker 2 values are also in reference to the delta
marker.
MARKER
All Off Marker Search Max Search
Mkr
−
> Max
Prior Menu Prior Menu Marker Functions
Delta Mkr ON
Prior Menu Marker Search Min Search Mkr
−
> Min
Prior Menu Prior Menu 2:
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...