Measuring S-Parameters using a Two-Port Calibration
Enter the Measurement Parameters
1. Press the following keys on the analyzer:
(
or
)
2. Press the softkey for any type of S-parameter measurement.
NOTE
This example measurement uses the default instrument parameters for
the S-parameter measurement you selected. If your particular
measurement requires different parameters (such as frequency range,
source power level, number of data points, and sweeptime), enter them
now.
Perform a User Two-Port Calibration
This example describes how to perform a two-port calibration. This
calibration uses known standards to correct for directivity, source match,
load match, and frequency response errors when using narrowband
detection. To correct for errors caused by crosstalk, refer to the definition
of a user two-port calibration in
.
When you perform a two-port calibration, the analyzer performs
correction at each data point across the selected frequency band. The
default number of data points per sweep is 201, but you can select any
number between 3 and 1601. Interpolation recalculates the error
correction array for reduced frequency spans. If the frequency span is
increased, the calibration is invalidated and the default two-port
calibration is automatically restored.
To perform a two-port calibration, you will need one of the following
calibration kits depending on the nominal impedance of your analyzer:
PRESET
User Preset
Factory Preset
MEAS 1
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...