3-44
ES User’s Guide
Making Measurements
Measuring S21 Forward Transmission using an Enhanced Response Calibration
Calibrating for Insertable Devices
When doing a through-standard calibration, normally the test ports
mate directly together. For example, two cables with the appropriate
connectors can be connected directly without a through adapter,
resulting in a zero-length through-standard. An insertable device is a
device that can be inserted in place of a zero-length through-standard. It
has the same type of connector on each port, but with a different sex, or
the same type of sexless connector on each port (APC-7, for example).
provides detail about when a user-defined calibration is
necessary, and information about other calibrations available for S
21
measurements. If you wish to perform an enhanced response calibration
on your instrument for an S
21
measurement, execute the following steps:
1. Press
.
2. The instrument prompts you to connect four standards—open, short,
load, and through cable—as shown in the following graphic.
3. Press
after connecting each standard.
CAL
User Response Enhanced Response
Open, Short, Load Connections
Through Cable Connection
Measure Standard
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...