ES User’s Guide
3-5
Making Measurements
Measuring Devices with Your Network Analyzer
Refer to
for the following discussion. The
network analyzer receiver has two signal detection modes:
• broadband detection mode
• narrowband detection mode
There are two internal broadband detectors: B* and R*. External
broadband detectors can also be used when connected to the X and Y
ports on the rear panel of the analyzer. When the network analyzer is in
the broadband detection mode, it measures the total power of all signals
present at these measurement ports, independent of signal frequency.
This enables the characterization of frequency translation devices such
as mixers, receivers, and tuners, where the RF input and output
frequencies are not the same.
labels the transmitted signal for
broadband detection input as B*, and the reference signal as R*.
When the network analyzer is in the narrowband detection mode, the
receiver is tuned to the source frequency. This technique provides greater
dynamic range by decreasing the receiver's bandwidth.
receiver B as the narrowband detection input for the forward
transmitted signal and the reverse reflected signal. Receiver A acts as
the narrowband detection input for the reverse transmitted signal and
the forward reflected signal. Receiver R is shown as the narrowband
detection input for the reference signal.
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...