ES User’s Guide
6-31
Calibrating for Increased Measurement Accuracy
Calibration Kits
How to download a calibration kit ASCII file. After creating a
calibration kit ASCII file, you can download the program by following
these steps:
NOTE
Calibration kit definitions must be in DOS format. LIF format is not
supported for cal kit definitions.
1. Insert the disk with the cal standard definitions into the analyzer's
built-in disk drive.
2. Press
and note the first word of the second line on
the screen. If it is not "INT:\" (for internal disk), press
.
3. Highlight the file (CALKIT, in the case of the Example Programs
Disk).
4. Press
and wait for the prompt:
Loaded cal kit information for 4 calibration standards
Recall of cal kit from CALKIT complete
5. Press
, and use the
front panel knob or the
keys to highlight the desired
port in the display’s table. Next, select the softkey that corresponds to
the User Cal Kit where the ASCII program stored the calibration kit
(default is User Cal Kit A).
NOTE
Cal kit coefficients are displayed in the cal kit block of the operating
parameters screen; press
;
press
a total of four times.
Step 3: Verify
Performance
Once a measurement calibration has been generated with a user-defined
calibration kit, its performance should be checked before making device
measurements. To check the accuracy that can be obtained with the
user-defined calibration kit, a device with a well-defined frequency
response (preferably unlike any of the standards used) should be
measured. The verification device must not be one of the calibration
standards: measurement of one of these standards is merely a measure
of repeatability.
To achieve more complete verification of a particular measurement
calibration, accurately known verification standards with a diverse
magnitude and phase response should be used. NIST-traceable
standards (such as HP verification standards) are recommended to
achieve verifiable measurement accuracy.
SAVE RECALL
Select Disk
Internal 3.5” Disk Prior Menu
Recall State
CAL
More Cal Cal Kit More Cal Kits
SYSTEM OPTIONS
Operating Parameters
Next Screen
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...