6-18
ES User’s Guide
Calibrating for Increased Measurement Accuracy
Calibration Choices
NOTE
When toggled on, the
softkey allows you to remove
systematic errors caused by crosstalk. To select it, press
. A two-port
calibration always measures the same parameters, regardless of the
setting of
. Therefore, no further measurements of
calibration standards are required when you enable the isolation
calibration feature. When toggled ON, this feature will change the
analyzer’s calculations before displaying the data. Crosstalk can be a
problem with high-isolation devices and high-dynamic range devices.
Since this isolation calibration adds noise to the error model, you should
only select it if it is really needed.
Transmission Calibration Interpolation
Widening the frequency span after performing any of these calibrations
will invalidate them, and restore the default calibration. You may
narrow the span and the analyzer will interpolate between calibration
points for the narrower span.
Note that after you have performed a calibration, a "C" appears in the
upper right hand corner of the display. This "C" indicates that a
user-defined cal is in use. If you change to a narrower span, note that the
"C" changes to "C?", indicating the analyzer is now interpolating between
calibrated measurement points. The "C?" notation also appears when
other system parameters, such as power, number of points, or sweep
time, have changed.
Isolation on OFF
CAL
User 2-Port Isolation on OFF User 2-Port
Isolation on OFF
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...