ES User’s Guide
9-53
Specifications
Options
75 Ohms (Option 1EC)
Provides 75 ohm system impedance.
Fault location and structural return loss (Option 100)
For fully characterizing cable performance and antenna-feedline
systems, this option provides both fault location and structural return
loss capability. Fault-location measurements help identify where
cable or system faults, such as bends, shorts, or corroded or damaged
connectors, occur. In addition to displaying faults in terms of distance
into the cable or feedline, the magnitude of the fault is also displayed.
Structural return loss is a special case of return loss (reflection)
measurements, optimized for measuring periodic reflections of small
magnitude. These periodic reflections can occur from physical damage
to the cable caused by rough handling, or from minor imperfections
imparted during the manufacturing process. Structural return loss
problems occur when these periodic reflections sum at
half-wavelength intervals, causing high signal reflection (and low
transmission) at the corresponding frequency.
Step Attenuator (Option 1E1): HP 8712ET/8714ET only
This option adds a built-in 60 dB step attenuator, extending the
source output power low-end range to
−
60 dBm. This attenuator is a
standard feature on the HP 8712ES/8714ES.
Transport case and fault location and structural return
loss (Option 101)
Combines a rugged transport and operation case (HP part number
08712-60059) with Option 100 for field measurements of fault
location and structural return loss.
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...