4-18
ES User’s Guide
Using Instrument Functions
Using Markers
To Use Marker Math Functions
The three marker math functions—statistics, flatness, and RF filter
stats—perform certain mathematical calculations on the amplitude data
of user-defined trace segments.
For measurement channel 1, the trace segment is defined with markers 1
and 2; for measurement channel 2, the trace segment is defined with
markers 3 and 4. The marker math parameters are updated after each
sweep, as well as any time a marker is moved. Regular marker tracking
is not available with the marker math functions.
NOTE
You cannot have marker math functions active at the same time as
marker search functions.
To Use Marker
Statistics
The marker statistics function measures a user-defined segment of the
measurement trace and calculates the following:
• frequency span
• mean amplitude
• standard deviation of the amplitude response
• peak-to-peak ripple
Limit testing may be performed on the statistical mean and peak-to-peak
ripple. See
“To Use Marker Limit Functions” on page 4-32
information.
1. On measurement channel 1, press
and set markers 1 and
2 to define the beginning and end of the trace segment that you want
to measure. (When using measurement channel 2, use markers 3 and
4 to define the trace segment.)
2. Press
.
3.
Figure 4-12, “Marker Statistics Function,”
segment. Notice the marker readout in the upper right corner of the
display.
MARKER
Marker Functions Marker Math Statistics
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...