6-14
ES User’s Guide
Calibrating for Increased Measurement Accuracy
Calibration Choices
To Perform a Normalization Calibration
Normalization is the simplest type of calibration. The analyzer stores
data into memory and divides subsequent measurements by the stored
data to remove frequency response errors.
Follow these general steps when performing a normalization calibration:
1. Set up the analyzer for your measurement:
• Select the type of measurement.
• Enter operating parameters other than the default.
2. Connect your equipment as you would for an actual measurement,
but omit the DUT.
3. Press
or
.
4. Notice that the top of the display on the analyzer now shows the type
of measurement followed by "/M" to indicate that the displayed data
is actually the measurement divided by memory.
CAUTION
The normalization cal will be invalidated if any frequency settings are
changed after calibration. There is no correction interpolation with a
normalization cal.
CAUTION
If you decide to perform any other type of calibration, be sure to turn off
normalization first by pressing
or
. Note that the "/M" indicator disappears when
normalization is turned off.
DISPLAY
Normalize
CAL
Normalize on OFF
DISPLAY
Data
CAL
Normalize ON off
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...