Configuring the Numerical Calculation Settings
Calculation type
Setting
Description
Example of screen
Time to level
Pulse count
Target channel
(Analog, logic, waveform
calculation)
Allows you to specify
calculation target
channels.
Level
*
Detects the time when
a waveform crossed
the level specified here
or a pulse count that
crossed the level. No
setting is available for
logic channels.
Slope
Detects the time when
a waveform crossed
the level specified here
in the positive direction
or a pulse count that
crossed the level in the
positive direction.
Detects the time when
a waveform crossed the
level specified here in
the negative direction
or a pulse count that
crossed the level in the
negative direction.
Filter(samples)
(
Off
,
10
to
10,000
)
Allows you to enter
duration used to
determine that a
waveform crosses the
specified level.
Only if a waveform does
not cross the level again
within the specified filter
duration after it has
crossed the level, the
waveform is considered
to have crossed the
level.
This is useful to
eliminate false level-
crossing events caused
by noise.
*: The range that can be set are between −9.9999E+29 and −1.0000E−29, 0, and b1.0000E−29 and
+9.9999E+29. A value to five significant figures can be set (a value to 10 significant figures can be set for a
time of the level at time setting).
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......