Configuring Measuring-Module-Specific Settings
Configuring Model 8967 Temp Unit settings
> [Channel] > [8967]
1
2
3
4
5
6
1
Click the
[Mode]
box, and then choose a thermocouple type from the list.
Choose an option depending on the type of a thermocouple to be used.
Mode
Measurable range
Mode
Measurable range
K
−200°C to 1350°C
R
0°C to 1700°C
J
−200°C to 1100°C
S
0°C to 1700°C
E
−200°C to 800°C
B
400°C to 1800°C
T
−200°C to 400°C
W
0°C to 2000°C
N
−200°C to 1300°C
2
Click the area that includes
[RJC]
.
The setting dialog box appears.
3
Click the
[RJC]
box, and then choose between
[Int.]
and
[Ext.]
from the list for reference
junction compensation.
Int.
Executes the reference junction compensation inside the module.
(Measurement accuracy: The sum of the accuracy of the temperature measurement
and that of the reference junction compensation)
Ext.
Does not execute the reference junction compensation inside the module.
(Measurement accuracy: The accuracy of the temperature measurement only)
When connecting a thermocouple directly to the module, choose
[Int.]
.
When connecting a thermocouple via a reference junction device that includes a zero-point bath, choose
[Ext.]
.
3
Advanced Functions
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......