71
Configuring Generator-Module-Specific Settings
Pattern reproducing frequency
Enter a clock frequency of pulse pattern. (In 10 Hz increments)
•
Settable clock period range: 0 to 0.1 seconds
•
You can freely specify the clock cycle as long as it falls into the above-mentioned range. The cycle of the
actually outputted waveform becomes the nearest value among the reciprocals of settable frequencies.
0
Hz to
120000
Hz
Output configuration
Choose an output status.
•
The ground potential is cross-channel and not insulated.
TTL
TTL-level output (amplitude: 0-5 V)
OC
Open-collector output
With the open-collector output setting
•
Limit the collector-emitter voltage to 50 V.
•
Output signals require a response time of about 5 µs or less to rise from 10% to 90%. (With a load capacity
of 1000 pF and a pull-up resistance of 1 k
Ω
, values for reference purposes)
Output
Switch waveform output settings.
The instrument may momentary output a high-level signal at the times of power-on and power-off.
OFF
Does not output a waveform
ON
Outputs a waveform
Confirmation of the setting
After finishing the setting, click
[Send setting]
.
The instrument confirms the settings and outputs a waveform.
3
Advanced Functions
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......