Configuring Generator-Module-Specific Settings
Offset
Allows you to set an offset of the output sine wave.
Accuracy of the output voltage, which consists of amplitude and an offset value, is guaranteed in
the range of −10 V to +10 V.
An output waveform will appear with its upper parts limited to about +14 V and lower about −14 V
when the sum of the amplitude and offset value is set to out of the accuracy-guaranteed range.
−10.000 V
to
+10.000 V
4
Choose whether to output a waveform.
Select a state at the time when no waveform is outputted
OFF (OPEN)
Does not output any waveform
Disconnect the output terminals from the internal circuit and short-circuits the terminal.
Internal
circuit
Output
terminal
OFF (SHORT)
Does not output any waveform
Disconnect the output terminals from the internal circuit and short-circuits the terminal.
Internal
circuit
About 20
Ω
Output
terminal
ON
Outputs a waveform
5
Click
[Send setting]
.
The instrument confirms the settings and outputs a waveform.
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......