Configuring the Pre-trigger settings
3
Click the
[
Trigger priority]
button to choose a way to handle triggers activated while the
instrument is filling the pre-trigger memory.
You can choose whether the instrument is triggered when the trigger condition is satisfied while the instrument
is filling the pre-trigger memory.
•
When the pre-trigger is enabled, the instrument is not triggered once the measurement has started until a
certain period has elapsed (while the instrument is filling the pre-trigger memory).
•
The message
[Waiting for Pre-Trigger]
appears on the screen.
Off
Ignores triggers while filling the pre-trigger memory.
On
Accepts a trigger while filling the pre-trigger memory.
When the trigger condition is satisfied while the message
[Waiting for Pre-Trigger]
is being
displayed
Example: When the pre-trigger length is set at 50%
Trigger priority:
[Off]
Ignores triggers.
Recording length
before the trigger
50% of recording length
after the trigger
[Waiting for Pre-Trigger]
[Waiting for
Trigger]
[Storing]
Starting a measurement
(After having internally recorded the 50%-recording-length -data,
displays
[Waiting for Trigger]
.)
Trigger priority:
[On]
When the instrument is triggered while it is filling the pre-trigger memory, the length
of the data actually recorded may be shorter than the specified recording length. (In
this case, the pre-trigger recording length is shortened. The recording length after a
trigger is calculated by subtracting the specified pre-trigger length from the specified
recording length.)
The instrument is triggered.
Recording length
before the trigger
50% of recording length
after the trigger
[Waiting for Pre-Trigger]
[Storing]
(The instrument is triggered while 50%-recording-length data is
being internally recorded)
Starting a measurement
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......