Configuring Generator-Module-Specific Settings
Configuring Model U8794 VIR Generator Unit settings
Choose
[11:U8794]
.
Notch frequency
Choose a power frequency according to the power frequency in your region.
An incorrect power frequency setting causes measured values to be unstable.
50
Allows you to set the notch frequency at 50 Hz.
60
Allows you to set the notch frequency at 60 Hz.
Mode
Allows you to choose an output type.
V
DC voltage
I
DC current
R
Resistance (simulated)
AC
AC voltage
Specifying voltage values
Enter an output DC current value.
−0.1 V
to
5.3 V
The output current range of −5 mA to +5 mA.
CAUTION
The instrument accepts a load resistance of 1 k
Ω
or more. Do not connect a load
resistance lower than the permissible value or short-circuit an output terminal. Damage
to the instrument could result.