59
Configuring Measuring-Module-Specific Settings
Configuring Model U8974 High Voltage Unit settings
> [Channel] > [U8974]
3
1
2
1
In the
[Mode]
area, click the
[DC]
or
[RMS]
to choose a measurement mode.
DC
For voltage measurement
RMS
For RMS measurement
2
(When you have changed the measurement mode) Click
[Zero adjust]
.
The instrument performs zero-adjustment. Execute zero-adjustment without any input.
3
Click the area that includes
[Response]
.
The setting dialog box appears.
4
Click the
[Response
]
box, and then from the list, choose a response time for RMS
measurement.
Fast
Sets the response time at 150 ms.
Normal
Sets the response time at 500 ms.
Slow
Sets the response time at 2.5 seconds.
Setting the response time to
[Slow]
can stabilize measured values if frequency is relatively low or voltage
fluctuates severely.
5
Click
[Close]
.
The setting dialog box closes.
3
Advanced Functions
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......