Configuring Generator-Module-Specific Settings
Voltage generation and measurement function
Outputting an AC voltage
The instrument can output a sine wave with a DC voltage superimposed.
DC offset voltage
0.1 V to 2.5 V
Amplitude
0.1 V p-p to 5.0 V p-p
Frequency
10 Hz, 20 Hz, 50 Hz, 100 Hz
CAUTION
The instrument accepts a load resistance of 1 k
Ω
or more. Do not connect a load
resistance lower than the permissible value or short-circuit an output terminal. Damage
to the instrument could result.
IMPORTANT
If you choose a frequency other than the frequencies that can be specified, the instrument sets a
frequency to the one that is the closest.
Output
Switch each output on and off.
OFF
Does not output any waveform.
ON
Outputs a waveform.
Switch
You can switch between the short-circuited state and open state for the OUTPUT terminal.
Open
Disconnects the OUTPUT terminal from the output circuit and the GND terminal.
Short
Connects the OUTPUT terminal with the GND terminal.
Normal
Connects the OUTPUT terminal with the internal output circuit.
Confirmation of the setting
After finishing the setting, click
[Apply]
.
The instrument confirms the settings and outputs a waveform.
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......