![Hioki MR8740-50 Instruction Manual Download Page 177](http://html1.mh-extra.com/html/hioki/mr8740-50/mr8740-50_instruction-manual_2137326177.webp)
171
Operators of Waveform Calculation and Calculation Results
b
i
:
i
th data point of calculation results,
d
i
:
i
th data point acquired across the source channel
Waveform calculation
type
Description
1st-order integration
(INT)
2nd-order integration
(INT2)
To calculate values of 1st-order and 2nd-order integration, the instrument uses the
trapezoidal formula.
The instrument integrates data
d
1
to
d
n
considering them as the corresponding data for
the sampling time
t
1
to
t
n
.
Arithmetic expressions of 1st-order integration
Point
t
1
I
1
= 0
Point
t
2
I
2
= (
d
1
+
d
2
)
h
/ 2
Point
t
3
I
3
= (
d
1
+
d
2
)
h
/ 2 + (
d
2
+
d
3
)
h
/ 2 =
I
2
+ (
d
2
+
d
3
)
h
/ 2
↓
Point
t
n
I
n
=
I
n-1
+ (
d
n-1
+
d
n
)
h
/ 2
I
1
through
I
n
: Calculation result data
h
=
Δ
t
: Sampling period
Arithmetic expressions of 2nd-order integration
Point
t
1
II
1
= 0
Point
t
2
II
2
= (
I
1
+
I
2
)
h
/ 2
Point
t
3
II
3
= (
I
1
+
I
2
)
h
/ 2 + (
I
2
+
I
3
)
h
/ 2 =
II
2
+ (
I
2
+
I
3
)
h
/ 2
↓
Point
t
n
II
n
=
II
n−1
+ (
I
n−1
+
I
n
)
h
/ 2
II
1
through
II
n
: Calculation result data
Note Integration is susceptible to minor deviation of the zero-position. Be sure to
execute zero-adjustment before performing a measurement. In addition, zero-
point offset correction may be required.
Example: To adjust the zero point by 0.124 mV
Z1 = INT (CH1-0.000124)
Waveform shifting
by PLC delay time
specified in digital
voltmeter module
(PLCS)
The instrument shifts a waveform by a frequency (PLC) specified in a digital voltmeter
module and a delay time of PLCS.
Since digital voltmeter modules calculate averages during periods specified in the
NPLC setting, observed waveforms will lag behind waveforms acquired with Model
8966 Analog Unit by a half of the NPLC setting.
The PLCS calculation advances the waveforms acquired with the digital voltmeter
module by the delay time and thus compensates the results for the lag behind the
waveforms acquired with Model 8966.
Note If the end of the calculation result contains no data, this part measures a voltage
of 0 V.
Average value (PAVE)
Calculates an average value of waveform data to use this numerical calculation result
for waveform calculations.
1
𝑖𝑖
=
1
∑
=
AVE
n
n
d
𝑖𝑖
AVE
: Average
n
: Number of data points
d
i
:
i
th data point acquired across
the channel
Maximum value
(PMAX)
Calculates the maximum value of waveform
data to use this numerical calculation result
for waveform calculations.
Maximum
Minimum value (PMIN)
Calculates the minimum value of waveform
data to use this numerical calculation result
for waveform calculations.
Minimum
8
W
aveform Calculation Function
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......