174
Configuring the memory division settings
9.1
Configuring the memory division settings
> [Status] > [Condition]
1 2
3
4
1
Click the
[Memory divide]
button to set it to
[On]
.
2
Click the
[Division]
box, and then enter the number of blocks the memory capacity is divided
into.
2
,
4
,
8
,
16
,
32
,
64
,
128
,
256
,
512
,
1024
3
Click the
[Points]
box, and then enter a recording length.
The maximum recording length varies depending on the division number specified in
[Division]
.
4
Enter the start block number in
[Start block]
.
The instrument stores the waveforms in the block with the number you specified.