137
Configuring the Numerical Calculation Settings
3
Click the
[Calculation Type]
box, and then choose a calculation type from the list.
Off
The instrument does not perform any calculation.
Average value
Average value of waveform data
RMS
RMS value of waveform data
P-P
Peak-to-peak value of waveform data
Maximum
Maximum value of waveform data
Time to maximum
Period of time elapsed from a trigger point to the time of the maximum value
Minimum
Minimum value of waveform data
Time to minimum
Period of time elapsed from a trigger point to the time of the minimum value
Period
*
2
Period of waveform data
Frequency
*
2
Frequency of waveform data
Rise time
*
1
Rise time of waveform data
Fall time
*
1
Fall time of waveform data
Standard deviation
Standard deviation of waveform data
Area
Area enclosed by the horizontal axis and waveform data
X-Y area
Area enclosed by an X-Y composite curve
Time to level
*
2
Period of time elapsed from a trigger point to the time when waveform data
reached the specified level
Level at time
*
2
Measured value when the specified time has elapsed from a trigger point
Pulse width
*
2
Pulse width of waveform data
Duty ratio
*
2
Duty ratio of waveform data
Pulse count
*
2
Pulse count of waveform data
Arithmetic operations
Four arithmetic operations of numerical calculation results
Time difference
*
2
Time lag between phenomena A and B
Phase contrast
*
2
Time lag between phenomena A and B represented as a phase difference
High level
*
1
High level of waveform data
Low level
*
1
Low level of waveform data
Intermediate value
Intermediate value of waveform data
Amplitude
*
1
Amplitude value of waveform data
Overshoot
*
1
Overshoot value of waveform data
Undershoot
*
1
Undershoot value of waveform data
+Width
*
1
Time value above the intermediate level
−Width
*
1
Time value below the intermediate level
Burst width
*
2
Burst width of waveform data
Accumulation
Accumulated value of waveform data
Angle of XY waveform
Slope angle of a regression line on an X-Y composite curve
*1: Unable to make any calculations using data acquired across the measurement channels of Model MR8990
and Model MR8990 Digital Voltmeter Unit.
*2: The settings can also be configured for the logic channels.
7
Numerical Calculation Function
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......