79
Configuring Generator-Module-Specific Settings
To perform precise measurement
IMPORTANT
Be careful of the following points when the instrument outputs a current of 100 µA or less with
the current output function, when it generates a resistance of 50 k
Ω
or more with the resistance
generation function, or when a device connected with the instrument has an output resistance of
50 k
Ω
.
•
We recommend you to use measurement cables with an insulating sheath made of
polyethylene (PE) or polytetrafluoroethylene (TFE). Using measurement cables with a low-
resistance insulating sheath will increase leak currents, with the result that accuracy is
significantly affected.
•
The instrument will become susceptible to exogenous noise. We recommend you to use
shielded measurement cables or connect a capacitor between the OUTPUT terminal and GND
terminal.
3
Advanced Functions