116
Triggering the Instrument at Regular Intervals (Interval Trigger)
5.8
Triggering the Instrument at Regular Intervals
(Interval Trigger)
Start triggers can be activated at specified intervals. Setting the recording mode to
[Repeat]
allows
the instrument to record waveforms at regular intervals.
•
When using the pre-trigger, the instrument starts monitoring interval-trigger times after the first
pre-trigger time elapses since the start of measurement.
•
No start triggers are activated by any interval triggers while the instrument is filling the pre-trigger
memory. An interval trigger triggers the instrument while the instrument is waiting for a trigger
after the instrument has filled the pre-trigger memory.
•
Since the clock is internally corrected, displayed trigger times may not synchronize with the
intervals of the interval trigger.
> [Trigger] > [Common]
2
3
1
1
Click
[Interval trigger]
.
The interval trigger setting dialog box will appear.
2
Click
[Interval trigger]
to set it to
[On]
.
3
Click the
[h]
,
[min]
, and
[s]
boxes, and then enter a period of time for interval triggers.
An internal trigger is generated at the start of measurement, and triggers are repeatedly generated at the
specified intervals.
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......