![Hioki MR8740-50 Instruction Manual Download Page 18](http://html1.mh-extra.com/html/hioki/mr8740-50/mr8740-50_instruction-manual_2137326018.webp)
12
Configuring the Input Channel settings
3
Click the
[
Range (f.s.)]
box, and then choose a measurement range from the list.
Select a measurement range for each channel. The value of the range represents its maximum displayable
value (f.s.).
See the following table for the full-scale resolution of each module.
If the input voltage exceeds the measurable range (overrange occurs), change the measurement range to one
with a lower sensitivity.
For Model MR8990 and Model U8991, starting a measurement using a high-sensitivity range with the input
terminal open causes an input signal to be overrange.
After changing the measurement range, check values of the level, upper limit, lower limit, and other values of
the trigger, search, and numerical calculation functions.
Module
Resolution (LSB)
Model 8966 Analog Unit
Model 8971 Current Unit
Model 8972 DC/RMS Unit
2,000
Model 8967 Temp Unit*
20,000
Model 8968 High Resolution Unit
Model U8974 High Voltage Unit
Model U8975 4ch Analog Unit
Model U8977 3CH Current Unit
Model U8978 4CH Analog Unit
32,000
Model U8969 Strain Unit
Model U8979 Charge Unit
25,000
Model 8970 Freq Unit (Power frequency mode)
2,000
Model 8970 Freq Unit (Count mode)
40,000
Model 8970 Freq Unit
(Frequency mode, rotation speed mode, duty ratio mode, pulse width mode)
10,000
Model MR8990 Digital Voltmeter Unit
Model U8991 Digital Voltmeter Unit
1,000,000
*: For the Model 8967 Temp Unit, the valid range varies depending on the thermocouples. For more
information about resolution, refer to “Model 8967 Temp Unit” in “5.2 Specifications of the Options” in Quick
Start Manual.
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......