Configuring Measuring-Module-Specific Settings
Configuring Model 8971 Current Unit settings
> [Channel] > [8971]
4
1, 2
3
1
Confirm the output rate displayed in the
[Mode]
area.
The instrument automatically recognizes a current sensor connected to Model 8971 Current Unit and displays
it as follows:
20A/2V
When one of the following current sensors is connected: Model 9272-10 (20 A range)
and Model CT6841.
200A/2V
When one of the following current sensors is connected: Model 9272-10 (200 A range),
Model CT6843, and CT6863.
50A/2V
When Model CT6862 AC/DC Current Sensor is connected
500A/2V
When one of the following current sensors is connected: Models 9709, CT6844,
CT6845, CT6846*, and CT6865*.
None
When no current sensor is connected.
IMPORTANT
*: When Model CT6846 or Model CT6865 connects to Model 8971 Current Unit via Model 9318
Conversion Cable, the instrument recognizes the sensor as a 500 A AC/DC sensor. Set the
conversion ratio at 2.00 in the scaling setting to obtain correct current values.
2
In the
[Mode]
area, click the
[DC]
or
[RMS]
to choose a measurement mode.
DC
For current measurement
RMS
For RMS measurement
3
(When you have changed the measurement mode) Click
[Zero adjust]
.
The instrument performs zero-adjustment. Execute zero-adjustment without any input.
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......