77
Configuring Generator-Module-Specific Settings
Switching OUTPUT terminals
Terminal setting
Internal circuit
Open
Output
circuit
OUTPUT
GND
Short
Output
circuit
OUTPUT
GND
Normal
Output
circuit
OUTPUT
GND
IMPORTANT
Do not set the OUTPUT terminal status to
[Short]
when a power supply device connects with the
OUTPUT terminal. A short-circuit current will flow, resulting in damage to the instrument or power
supply device.
Self-diagnosis function
Simple test
At the time when you send a query for testing, the instrument measures an output value of the
instrument with the internal measuring circuit and return the specified value and a measured value.
For the voltage output setting, the instrument sends back the specified voltage and a measured
voltage value.
For the current output setting, the instrument sends back the specified current and a measured
current value.
3
Advanced Functions