49
Configuring Measuring-Module-Specific Settings
Configuring Model U8969 Strain Unit settings
Model U8969 Strain Unit can execute auto-balance.
Executing auto-balance regulates the reference output level of a transducer at the specified zero
position. Auto-balance is available for Model U8969 Strain Unit only.
You can use Model 8969 Strain Unit you own with this instrument. The instrument displays the
model name of Model 8969 Strain Unit as
[U8969]
.
Before executing auto-balance
•
Turn on the instrument and leave it for 30 minutes to allow the internal temperature of the module
to stabilize.
•
After connecting a strain gauge converter to the module, execute auto-balance without any input
including distortion.
•
You cannot execute auto-balance during measurement.
•
No operation is accepted during auto-balance.
To execute auto-balance on the channel screen of each channel
> [Channel] > [U8969]
Click
[Auto balance]
, and then choose an auto-balance setting from the list.
One channel only
Executes auto-balance for only a channel displayed on the channel screen.
All
Executes auto-balance for all of the channels Model U8969 is installed in.
3
Advanced Functions
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......