56
Configuring Measuring-Module-Specific Settings
Configuring Model 8972 DC/RMS Unit settings
> [Channel] > [8972]
3
1
2
1
In the
[Mode]
area, click the
[DC]
or
[RMS]
to choose a measurement mode.
DC
For voltage measurement
RMS
For RMS measurement
2
(When you have changed the measurement mode) Click
[Zero adjust]
.
The instrument performs zero-adjustment. Execute zero-adjustment without any input.
3
Click the area that includes
[Response]
.
The setting dialog box appears.
4
Click the
[Response
]
box, and then from the list, choose a response time for RMS
measurement.
Fast
Sets the response time to about 100 ms.
Normal
Sets the response time to about 800 ms.
Slow
Sets the response time to about 5 seconds.
Usually, use
[Fast]
. Setting the response time to
[Normal]
or
[Slow]
can stabilize the measured values if the
frequency is relatively low or the voltage fluctuates severely.
5
Click
[Close]
.
The setting dialog box closes.
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......