5.9 Application Settings
229
5
Chapte
r 5
ANAL
YZER Functi
o
n
The higher the measurement frequency, the shorter the time that INDEX and EOM are high (off). When the
high (off) time is too short due to characteristics of the input circuit, the instrument can be configured to main-
tain the low (on) state for a preset time once EOM changes to low (on) before reverting the signal to high (off)
after the completion of measurement. The INDEX output method can be changed in the same manner.
See
"Chapter 11 External Control" (p. 363)
5.9.6
Setting the EOM Output Method
4
Procedure
Application Settings
1
Analyzer Initial Screen
Press
.
2
Application Settings
3
Setting the EOM output method
Press
to close the setting screen.
Setting the output method.
Settable range: 0.0001 to 0.9999 s
For HOLD and PULSE timing charts, see "Chapter
11 External Control" (p. 363).
Use
or
to set the EOM output time
for the PULSE setting.
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......