11.1 External Input/Output Connector and Signals
366
19
OUT OUT_OF_BINS
BIN judgment results
CIRCUIT_NG
Equivalent circuit analysis comparator
judgment result output (output when log-
ical AND of judgment results is FAIL)
Neg Level
20
-
(Unused)
-
(Unused)
-
-
-
21
IN
CALIB
DC adjustment request (Unused)
-
Neg
Level
22
IN
LD0
Select panel number
LD0
Select panel number
Neg
Level
23
IN
LD2
Select panel number
LD2
Select panel number
Neg
Level
24
IN
LD4
Select panel number
LD4
Select panel number
Neg
Level
25
IN
LD6
Select panel number
LD6
Select panel number
Neg
Level
26
IN
LD_VALID
Execute panel load
LD_VALID
Execute panel load
Neg
Level
27
-
ISO_COM
Isolated common sig-
nal ground
ISO_COM
Isolated common signal ground
-
-
28
OUT EOM
End of measurement
EOM
End of measurement
Neg
Edge
29
OUT INDEX
Analog measurement
finished
INDEX
Analog measurement finished
Neg
Edge
30
OUT BIN2, PARA1-IN
BIN judgment results
Comparator judgment
results of the first pa-
rameter
PARA1-IN,
PARA1_LMAX_IN
Analyzer comparator results of the first
parameter in AREA mode
(Output if all judgment results are IN)
Analyzer comparator results for the local
maximum value of the first parameter in
PEAK mode
(Output if PEAK is IN)
Neg
Level
31
OUT BIN4, PARA3-HI
BIN judgment results
Comparator judgment
results of the third pa-
rameter
PARA2-HI,
PARA2_LMAX_MEASNG
Analyzer comparator results of the sec-
ond parameter in AREA mode
(Output if even one HI judgment)
Analyzer comparator results for the local
maximum value of the second parameter
in PEAK mode
(Output if the vertical axis (measurement
value) is out of the range or there is no
comparison peak)
Neg
Level
32
OUT BIN6, PARA3-LO
BIN judgment results
Comparator judgment
results of the third pa-
rameter
PARA2-LO,
PARA2_LMAX_CONDNG
Analyzer comparator results of the second
parameter in AREA mode
(Output if even one LO judgment)
Analyzer comparator results for the local
maximum value of the second parameter in
PEAK mode
(Output if the horizontal axis (sweep setting)
is out of the range or there is no comparison
peak)
Neg
Level
33
OUT BIN8
BIN judgment results
PARA1_LMIN_MEASNG
Analyzer comparator results for the local
minimum value of the first parameter in
PEAK mode
(Output if the vertical axis (measurement
value) is out of the range or there is no
comparison peak)
Neg
Level
34
OUT BIN10
BIN judgment results
PARA1_LMIN_CONDNG
Analyzer comparator results for the local
minimum value of the first parameter in
PEAK mode
(Output if the horizontal axis (sweep setting)
is out of the range or there is no comparison
peak)
Neg
Level
35
OUT (Unused)
-
PARA2_LMIN_IN
Analyzer comparator results for the local
minimum value of the second parameter
in PEAK mode
(Output if PEAK is IN)
Neg
Level
36
-
(Unused)
-
(Unused)
-
-
-
37
-
(Unused)
-
(Unused)
-
-
-
Pin
I/O
LCR Mode
ANALYZER Mode
Logic
Signal name
Function
Signal name
Function
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......