11.2 Timing Chart
375
11
Chapte
r
1
1 External
Co
ntro
l
In continuous measurement mode, pressing
on the screen asking whether a trigger signal will be
input from EXT I/O will cause the judgment result to be output from the EXT I/O comparator result output sig-
nal lines after measurement of the panel set to be executed on the screen completes. The following provides
an example of measurement timing with the
trigger setting.
(In this timing example, the TRIG signal active edge is set to the falling edge (ON).)
11.2.4 Continuous Measurement (Step Operation)
Signal line
Description
INDEX
For both INDEX, a transition to HIGH is performed when the first panel measurement starts after the trigger
signal is input, and a transition to LOW is performed after measurement of the last panel is finished and the
judgment result has been output.
(The HIGH level is maintained during continuous measurement.)
EOM
EOM transitions to HIGH at the start of measurement after trigger signal input and to LOW each time mea-
surement completes.
Comparator result output
The judgment result and ERR vary according to the measurement state each time measurement completes.
*1: Reset at the same time as analog measurement starts.: HIGH
Not reset at the same time as analog measurement starts.: Last judgment result remains
(Start-of-measurement signal)
(Analog end of measurement signal)
(End of measurement signal)
(Comparator result output)
Last judgment
result
Judgment result
Judgment result
(Error output)
•
Panel load signals (LD-VALID, LD0 to LD6) cannot be used on the continuous measurement
screen.
See:
"Chapter 6 Continuous Measurement Function" (p. 265)
•
Comparator judgment results are output just as they are during measurement in LCR or analyzer
mode.
•
Whether the judgment results of comparator measurement are reset at the time of the mea-
surement start signal or updated at the point in time when measurement ends can be selected
on the instrument or by a communication command.
See:
"Setting the Delay Time from the Output of Comparator and BIN Judgment Results until Output of EOM
(LOW) and Resetting Judgment Results" (p. 115),(p. 226)
Description of communications commands on the included CD
(
:IO:RESult:RESet
)
•
For other timing chart times, refer to
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......