4.2 Setting Basic Settings of Measurement Conditions
48
Testing some types of sample is not possible using constant current. In this case, the following
symbol appears on the display:
In such a case, constant current testing is not performed. Change the constant current level
to a value not more than the value being shown as Imoni.
Reference When a 1 mH impedance is measured at 1 kHz, the CC operation range can be
obtained as follows.
Sample impedance
Zm
becomes as follows:
The impedance
Zm'
observed from the generator is as follows:
Accordingly, the current
Im
across both leads of the sample is as follows:
Because the generator output voltage range is 5[mV] to 5[V] for 1 kHz, the CC operation
range per the above expression is
Im
= 8.47[
A] to 42.3[mA].
In low Z high accuracy mode, the output resistance
Ro
becomes 10[
].
Xm
2
fL
=
Zm
=
Rm
+
jXm
=
0
[
] -
j62.8
[
]
Ro
: output resistance (100 [
])
Zm'
=
Ro
+
Zm
=
100
[
] -
j62.8
[
]
Im
Vo
Zm
------------
Vo
118.1
------------------
=
=
Vo
: generator output
|
Zm'
|
Vo
118.1
[
]
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......