4.4 Judging Measurement Results
90
The set reference value and upper and lower limit values are common to percentage mode
and percentage deviation mode.
• The actual internal operation calculates the lower limit comparison value with the following equation, and when
a value that is lower than the reference value is set, the minus (-) sign is required for the percentage setting
value.
Lower limit comparison value=reference value+ |reference value| ×
100
Percentage set value
• Settable range: -999.9999% to 999.9999%
8
Press
to close the setting screen.
5
• Settable range: -999.9999% to 999.9999%
6
Press
to confirm the upper limit value.
Press
and use the numeric keypad to
set the upper limit value.
Set the upper limit value as a percentage relative
to the reference value.
• The actual internal operation consists of calculating the upper-limit value of comparison using the equation
given below, and comparing it to the measurement value to enable a decision to be made.
Upper limit comparison value = reference value + |reference value| ×
7
Return to step
2
, press
, use the numeric keypad to enter the lower limit value, and
press .
Upper Limit Value Setting
When you do not want to set the upper
limit:
Press
.
100
Percentage set value
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......