7.1 Setting Open Circuit Compensation
275
7
Chap
ter 7 Erro
r Com
pensatio
n
All Compensation
5
6
Compensation starts.
Compensation value acquisition time:
Approximately 2 minutes
When you want to cancel compensation:
Press .
Compensation is cancelled and the compensation
screen is redisplayed.
(The open circuit compensation value state from last
time remains.)
When compensation failed:
7
Press
to close the setting screen.
• If compensation ends normally, the conductance
and susceptance are displayed.
• Compensation can be performed for impedances of at
least 1 k
.
Performing Open Compensation
You can use
and
to check the Conduc-
tance, susceptance of each compensation point.
4
All Compensation
Indicates the
compensation
numbers.
The compensation values from last time are dis-
played in a confirmation screen.
(If compensation has never been performed, the
compensation values become 0.)
Check that the measurement cable is in an open circuit
state.
When you do not want to acquire the compen-
sation values:
Press .
The setting screen is redisplayed, and the compensation
values from last time become valid.
Press .
Indicates the
measurement
frequencies.
Indicates the compensation results.
(Conductance, susceptance)
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......