14.2 Troubleshooting
406
Perform a system reset.
This will return all settings to their factory defaults.
See
Performing a full reset will restore all of the settings to the factory default settings.
Only perform a full reset in the following cases.
• When the normal reset screen cannot be displayed because of a problem with the instrument.
(After the full reset, perform a self check to confirm that there are no problems. (p. 307))
• When you have forgotten the passcode for the key lock.
AUTO ranging is un-
able to determine a
range.
Is a high-impedance element which is influenced by
noise being measured?
Use guarding.
See (p.A3)
Is an element of other than a capacitor being mea-
sured using the DC bias function?
Turn the DC bias function OFF.
See (p. 57)
A contact error is
generated even
though the connec-
tions are correct.
Is an element of other than a capacitor being mea-
sured using the DC bias function?
Turn the DC bias function OFF.
See (p. 57)
An error beep sound
is emitted continu-
ously.
Is the measurement value automatic output function
enabled?
When the measurement value automatic out-
put function is enabled, a send error occurs on
the measuring instrument side if the receive
operation is not performed on the PC side, and
a send error sound is emitted continuously
when, for example, there is an internal trigger.
Perform the receive operation on the PC side
and then perform measurement on the mea-
suring instrument side, or disable the measure-
ment value automatic output function.
See Description of communications commands
on the included CD
Symptom
Check Item, or Cause
Remedy and Reference
When no apparent cause can be established
Full Reset Procedure
1
Turn ON the switch on the back panel.
2
While the opening screen is displayed, press
the top right of the screen continuously.
Connect the power cable.
3
If the instrument still does not operate normally after the full reset, it needs to be repaired.
Contact your dealer, or a Hioki representative if you are not sure where the instrument was purchased.
The full reset is complete when a beeping sound
is emitted.
4
Opening Screen
After the full reset, the panel calibration screen is
displayed automatically. (p. 308)
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......