11.2 Timing Chart
372
In analyzer mode, if a trigger signal is input from the EXT I/O or
is pressed in the screen, the
judgment results are output from the signal line for comparator result output of the EXT I/O.
Furthermore, if the panel number is selected with the panel load signal when a trigger signal is input from the
EXT I/O, the measurement condition of that panel number is loaded and then measurement is performed.
The following provides an example of measurement timing when the trigger setting is
or
.
(In the timing examples, the valid edge of the TRIG signal is set to falling (ON).)
• If the trigger setting is set to STEP, INDEX and EOM transition to LOW every time the measurement for
each point ends, and then transition to HIGH when there is trigger input.
ERR also transitions to LOW each time measurement completes if a measurement error occurs.
• Whether the judgment results of comparator measurement are reset at the time of the measurement
start signal or updated at the point in time when measurement ends can be selected on the instrument
or by a communication command.
See:
"5.9.4 Setting the Delay Time from the Output of Comparator Judgment Results until Output of EOM (LOW) and
Resetting Judgment Results" (p. 226)
Description of communications commands on the included CD
(
:IO:RESult:RESet
)
• For other timing chart times, refer to "11.2.1 LCR Measurement" (p. 369).
11.2.2 Analyzer Measurement
*1: Reset at the same time as analog measurement starts.: HIGH
Not reset at the same time as analog measurement starts.: Last judgment result remains
(Start-of-measurement signal)
(Analog end of measurement signal)
(End of measurement signal)
(Comparator result output)
Judgment result
(Error output)
Last judgment
result
Signal line
Description
INDEX
The transition to HIGH is performed when measurement of the first sweep point starts after trigger signal input
and the transition to LOW is performed when the analog measurement of the last sweep point ends.
(The HIGH level is maintained during sweep measurement.)
EOM
The transition to HIGH is performed when measurement of the first sweep point starts after trigger signal input.
Measurement of the last sweep point ends and the transition to LOW is performed after judgment result output.
(The HIGH level is maintained during sweep measurement.)
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......