![Hioki IM3570 Instruction Manual Download Page 94](http://html1.mh-extra.com/html/hioki/im3570/im3570_instruction-manual_2137305094.webp)
4.4 Judging Measurement Results
86
The comparator measurement allows you to do the following.
The comparator decision mode can be set as one of the following:
4.4.1 Judging with Upper and Lower Limit Values
(Comparator Measurement Mode)
• Preset a reference value and upper and lower limit values as the judgment reference, and display a
judgment result as
HI
(higher than the upper limit value),
IN
(within the range set for the upper and
lower limit values), or
LO
(lower than the lower limit value).
• Output the judgment results to an external device (via the EXT I/O connector).
• Select different settings and perform judgment for up to two parameters.
• Be notified of judgment results by buzzer.
See
"4.5.13 Setting Operation Sounds (Beep Sounds)" (p. 128)
• Confirm the judgment result from the judgment result
indication LEDs on the front panel of the instrument.
See
"Judgment Result Indication LEDs" (p. 10)
(Front panel LEDs)
Lights green
Lights red
(When the compar-
ator measurement
result is HI or LO)
(When the comparator
measurement result is IN)
Measured value is above
upper limit
Pass (meets criteria)
Measured value is below
lower limit
Deviation Percent (
%) Setting (p. 91)
HI
IN
IN
LO
upper limit value[
%]
reference value[
]
lower limit value[
%]
Enter reference values and then set percentages corre-
sponding to the reference values as the upper limit and
lower limit values.
The measurement values are displayed in deviations (
%) from the refer-
ence value.
Percent (%) Setting (p. 89)
HI
IN
IN
LO
upper limit value[%]
reference value[
]
lower limit value[%]
Enter reference values and then set percentages corre-
sponding to the reference values as the upper limit and
lower limit values.
The measurement values displayed are the same as those of the mea-
surement parameters.
Absolute value (ABS) setting (p. 88)
HI
IN
LO
upper limit value
lower limit value
Set absolute values for the upper limit and lower limit values of
the measurement parameters.
The measurement values displayed are the same as those of the mea-
surement parameters.
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......