11.2 Timing Chart
370
Timing Chart Interval Descriptions
*1: There is an approximate error of 100
s in the delay time entered for Judgement Result
EOM for the
setting value. t1 is the reference value for when the setting value is 0.0000 s.
*2: t2 is the reference value for when trigger input for during measurement is disabled.(p. 117), (p. 228)
*3: • When the panel number is read by the panel load function, the response time is as shown in the table
below.
• When the trigger synchronous output function and trigger delay is enabled, wait times are included.
*4: Reference value for Measurement frequency: 100 kHz, Measurement speed: FAST,
Measurement range: HOLD (p. 399)
*5: With the conditions shown in the table below, the time is added.
• Since the speed of the rise (LOW
HIGH) of the comparator/BIN judgment result differs depending on the
configuration of the circuit connected to the EXT I/O, there is the likelihood of an incorrect judgment if the
level of the comparator/BIN judgment result acquired immediately after EOM output is used. To prevent this
from happening, a delay time (t1) between the comparator/BIN judgment result and the EOM can be set.
Furthermore, if the judgment result signal line of the EXT I/O is set to be reset simultaneously with the mea-
surement start signal, and a forced transition to the HIGH level is performed at the same time as TRIG, the
transition from LOW to HIGH does not occur when the judgment result is output after measurement ends.
As a result, the delay time between the judgment result and the EOM can be set to the minimum level. How-
ever, be careful because the judgment result confirmation interval is until the next trigger is accepted.
Furthermore, if the judgment result signal line of the EXT I/O is set to be reset simultaneously with the mea-
surement start signal, and a forced transition to the HIGH level is performed at the same time as TRIG, the
transition from LOW to HIGH does not occur when the judgment result is output after measurement ends.
As a result, the delay time between the judgement result output and EOM can be set to the minimum level.
As a result, the delay time between the judgement result output and EOM can be minimized. However, take
note that the evaluation result confirmation range is valid until the following triggers are accepted.
• During measurement, a trigger input from EXT /IO or communicating by interface may lead to a bigger dis-
persion of delay time between comparator or BIN judgement result output and EOM. As far as possible, try
not to control from external sources when carrying out measurement.
See
Resetting Judgment Results" (p. 115),(p. 226)
Description of communications commands on the included CD
(
:IO:OUTPut:DELay
,
:IO:RESult:RESet
)
Interval
Description
Time
(Approximate)
t1
From Comparator, BIN Judgement Result to EOM (LOW): Setting value for delay time
*
1
40
s
t2
From EOM width (LOW) to TRIG (LOW): Minimum time from end of measurement to next trigger
*
2, 5
400
s
t3
From TRIG(LOW) to INDEX (HIGH): Time from trigger to circuit response
*
3, 5
700
s
t4
INDEX width (HIGH): Minimum chuck time, switching chuck with INDEX (LOW) is possible
*
4
220
s
t5
EOM width (HIGH): Measurement time
*
4
600
s
t6
From TRIG width (LOW) to LD-VALID (HIGH) and CALIB (HIGH): Time to recognize panel number
t3
Measurement mode
Load mode
Response time
LCR
LCR+ADJ
12 ms
HARD
9 ms
ADJ
4 ms
ANALYZER
ANA+ADJ
120 ms
HARD
120 ms
ADJ
8 ms
Range
Level
Frequency
Low Z High Accuracy Mode
DC Bias
Time
100 m
, 1
All levels
DC, 4 Hz to 999.99 Hz
ON
OFF
2 ms
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......