4.4 Judging Measurement Results
87
4
Chapte
r 4
LCR Function
• The comparator judgment is made in the following order.
1.
If the measurement value is "OVER FLOW",
HI
is displayed.
(However, LO is displayed when the parameters are Y, Cs, Cp, G, and B.)
If the measurement value is "UNDER FLOW",
LO
is displayed.
(However, HI is displayed when the parameters are Y, Cs, Cp, G, and B.)
If the measurement value is “L NO CNTCT” or “H NO CNTCT”,
HI
is displayed.
2.
Whether the measurement value is higher than the lower limit value is judged, and
LO
is displayed if the judgment is NG.
3.
Whether the measurement value is lower than the upper limit value is judged, and
HI
is
displayed if the judgment is NG.
4.
When both 2 and 3 are satisfied,
IN
is displayed.
No test is performed to ensure that the upper limit value is greater than the lower
limit value, so no error message will be displayed if you set the upper limit value
and lower limit value the wrong way around.
• If the power is turned off while the comparator measurement screen is displayed, the
comparator measurement screen will be displayed when the instrument starts the next
time you turn the power on.
*
2: The following equation is used to calculate the
% value.
% =
|reference value|
measurement
×100
*
1: The following equation is used to calculate the comparison upper limit value and comparison lower limit value.
(In the case of the comparison lower limit value, if a value that is lower than the reference value is set, the minus
(-) sign is required for the percentage setting value.)
Upper limit comparison value (Lower limit comparison value) = reference value+ |reference value| ×
100
Percentage set value
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......