7.3 Compensating Values to Match Reference Values (Load Compensation)
289
7
Chap
ter 7 Erro
r Com
pensatio
n
• When all of the compensation conditions match for load compensation, compensation is
performed under the optimal conditions.
• If the current measurement frequency and compensation frequency do not match, an
error like the following is displayed in the initial screen.
• If a condition other than the compensation frequency does not match, compensation is
performed but a warning like the following is displayed in the initial screen.
• When the OPEN or SHORT compensation is valid, the load compensation is performed
for Z and
processed by the OPEN or SHORT compensation.
• In acquiring load compensation data (i.e., reference sample measurement), the OPEN/
SHORT compensation settings, that were defined before entry into the Load Compensa-
tion Screen, are valid.
• If the setting of the low Z high accuracy mode is changed, the compensation value
becomes invalid.
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......