4.2 Setting Basic Settings of Measurement Conditions
60
Depending on the measurement signal level, in some cases it is possible to damage the sample which is
being tested by applying to it a voltage or a current greater than its rated value.
For this reason, set a limit value to restrict the voltage that can be applied to the test sample or current that
can flow to the test sample.
4.2.7 Setting the Voltage/Current Limit
When open circuit voltage (V) or
constant voltage (CV) is set:
Set the current limit.
When constant current (CC) is set:
Set the voltage limit.
Procedure
LCR Initial Screen
• The present situation can be checked using the monitor display.
• The monitor display is different for V, CV, and CC.
First set the measurement signal level, and thereafter set the voltage or current limit.
The setting for voltage or current limit changes automatically to current or voltage limit,
according to the present measurement signal level setting.
See
"4.2.2 Setting the Measurement signal level" (p. 43)
1
LCR Basic Settings
3
When the measurement signal level is a
voltage (V, CV)
Current Limit Setting
When the measurement signal level is a
current (CC)
Voltage Limit Setting
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......