4.5 Setting Application Settings
105
4
Chapte
r 4
LCR Function
•
When the trigger synchronous output function is set to ON, there is a measurement time
delay because the unit enters a wait time which spans from when the measurement sig-
nal is output to when measurement starts.
"13.3 About Measurement Times and Measurement Speed" (p. 399)
•
When the trigger synchronous output function is set to ON, the set DC level may be out-
put if a measurement condition is changed. Also, output will stop when measurement is
performed once.
•
The measurement signal is output when the trigger signal is input and stops after mea-
surement ends.
•
In continuous measurement mode, the measurement signal stops after measurement of
the last panel ends.
3
Trigger Sync Setting
Select ON/ OFF for the trigger synchronous
output function.
Sets the trigger synchronous output function
to OFF.
Sets the trigger synchronous output function
to ON.
Use
or
to set the wait time from after
the measurement signal is output by applying
a trigger to the start of measurement.
5
4
Press
to close the setting screen.
Settable range: 0.0010 to 9.9999 s
When you want to return the time to the ini-
tial state:
Press .
The set time is set to 0.0010 s.
Trigger Sync Setting
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......