4.5 Setting Application Settings
113
4
Chapte
r 4
LCR Function
The instrument monitors the detectable range of the measurement signals (voltage and current) during mea-
surement, and displays a measurement error (OVERFLOW or UNDERFLOW) if that range is exceeded.
However, if measurement is performed under a noisy environment, the detection range may be exceeded
resulting in a measurement error regardless of whether an appropriate measurement range is set.
Under normal circumstances, the countermeasures described in "Appendix 4 Countermeasures Against
Incorporation of External Noise"(p. A5) need to be taken but if a measurement error occurs even after taking
the countermeasures, increase the setting value which defines the error of the measurement signal to reduce
the detection sensitivity for measurement errors so that you can reduce the frequency of measurement error
occurrence.
4.5.5
Setting the Detection Sensitivity for Measure-
ment Errors (Overflow or Underflow)
Procedure
LCR Initial Screen
Application Settings
1
Press
.
2
Application Settings
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......