7.2 Short Circuit Compensation
286
The compensation values from last time are dis-
played in a confirmation screen.
(If compensation has never been performed, the
compensation values become 0.)
5
Spot Compensation Value Setting
6
Compensation starts.
Compensation value acquisition time:
Approximately 5 seconds
Performing Short Compensation
Check that the measurement cable is in a short-circuit
state.
Press
.
When you want to cancel compensation:
Press .
Compensation is cancelled and the compensation
screen is redisplayed.
(The short circuit compensation values from last time
remain.)
When you do not want to acquire the compen-
sation values:
Press .
The setting screen is redisplayed, and the compensa-
tion values from last time become valid.
Spot Compensation
7
When compensation failed:
• If compensation ends normally, the effective resis-
tance and reactance are displayed.
• The possible compensation range is 1 k
or less for
impedance.
Use
or
to check the effective resistance
and reactance of each compensation point.
8
Press
to close the setting screen.
Indicates the
compensation
numbers.
Indicates the
measurement
frequencies.
Indicates the compensation results.
(Effective resistance, reactance)
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......