7.3 Compensating Values to Match Reference Values (Load Compensation)
294
Setting the Compensation Range
2.
Select the range for compensation.
The range that can be set varies depending on the com-
pensation frequency.
Compensation Range Setting
• To enable LOW Z: Press
of LOW Z.
• If you make a mistake during input:
press
to cancel the input and start again.
Compensation Range Setting
1.
Press
.
3.
Press
to close the setting screen.
Frequency
Settable
Range
Range Setting Screen
DC
Entire
range
4.00 Hz to
100.00 kHz
100.01 kHz to
1.0000 MHz
100 m
to
10 M
1.0001 MHz
to
5.0000 MHz
100 m
to
1 M
If the compensation frequency is not set, the compensation range cannot be set.
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......