4.2 Setting Basic Settings of Measurement Conditions
62
With the averaging function, the measured values can be averaged. Using this function, it is possible to
reduce fluctuations in the measured value display.
When the number of averaging times is 4, the number of measurements, measurement output points, and
measurement value calculation method during output are as follows.
4.2.8 Displaying Average Values (Averaging Set)
With internal trigger
A rolling average of the tested values over the set number of times for
averaging is always calculated backwards from the present.
(
When the sample to be tested is changed over, it takes a little time for a certain stabili-
zation time period until the results is reliable.)
With external trigger
An average of the test values is calculated over the set number of
times for averaging forwards from when the trigger is input.
M1
M2
M3
M4
M5
M6
M7
M8
M9
M10
M1
1
M1+M2
2
M1+M2+M3
3
M1+M2+M3+M4
4
M2+M3+M4+M5
4
M3+M4+M5+M6
4
M4+M5+M6+M7
4
M5+M6+M7+M8
4
M6+M7+M8+M9
4
M7+M8+M9+M10
4
Measurement points
Measurement value output
Moving average
Time
M1
M2
M3
M4
M5
M6
M7
M8
M9
M10
Time
Arithmetic average
M1+M2+M3+M4
4
M5+M6+M7+M8
4
Measurement points
Measurement value output
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......