4.3 Setting DC Resistance Measurement
76
4
Press
to close the setting screen.
3
DC Adjustment Setting
Select ON/ OFF for DC adjustment.
Obtains the offset value with the CALIB
signal of EXT I/O or a communication
command (:
DCResistance:ADJust
:DEMand
).
Obtains the offset value for each measure-
ment.
• Since the transient response time becomes long as a result of the impedance value of the measure-
ment object, set a long DC delay time.(p. 82)
When DC adjustment is ON, the measurement time becomes about twice as long because time is
required both for normal measurement and for offset measurement.
• DC adjustment measurement is performed in the following cases even if DC adjustment is set to OFF.
• When measurement is first performed after the power is turned on
• When the DC delay time is changed
• When the measurement range is changed
• When low Z high accuracy mode is turned ON/OFF
• When the measurement signal level is changed
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......